40
Annual Inspection
EN
nector).
7. Stop insufflation:
Press the
STOP
key.
The following values are displayed:
Actual pressure 0 mmHg
Gas consumption > 0.0 l
8. Press the
RESET
key.
Gas consumption 0.0 l
The basic function check of the device is complete.
10.3
Pressure Sensor Test
1. Select the
Standard Insufflation Mode
as the operating mode.
2. Do not press the
START/STOP
key.
CAUTION!
Never use the syringe to extract gas from the device.
3. Connect a manometer
(1)
and an air-filled syringe
(3)
to the insufflation
tube connection
(2)
.
4. Use the syringe to generate a pressure of at least 10 mmHg, which is indicat-
ed on the manometer.
Actual pressure display: 10 ±2 [mmHg]
5. Use the syringe to generate a pressure of at least 20 mmHg, which is indicat-
ed on the manometer.
Actual pressure display: 20 ±2 [mmHg]
6. Use the syringe to generate a pressure of at least 30 mmHg, which registers
on the manometer.
Actual pressure display: 30 ±2 [mmHg]
10.4
Pressure Monitoring Test
1. Select the
Standard Insufflation Mode
as the operating mode.
2. Select a nominal pressure of 15 mmHg and adjust in the user menu the gas
flow rate 1.
3. Use the syringe to generate a pressure of at least 19 mmHg, which registers
on the manometer.
An acoustic warning sound is emitted with a pressure of more than 19 mmHg
(1)
(3)
(2)
Содержание AirSeal iFS
Страница 1: ...EN Instructions for use...
Страница 4: ......
Страница 6: ...2 EN 15 AirSeal iFS Warranty 52 16 Test Log 53 16 1 Test Log 53 16 2 Return Form 54 Index 55...
Страница 20: ...16 Initial Device Startup EN House gas supply pressure OK House gas supply pressure too low...
Страница 57: ...Test Log 53 EN 16 Test Log 16 1 Test Log Date Results Comment Signature...
Страница 60: ...W55 186 EN AA...