Using and Controlling the AirSeal® iFS in its Different Modes
25
EN
seconds.
Increasing the nominal pressure to > 15 mmHg:
Status line depicts the message:
Safety threshold
:
Pressure >15 mmHg!
This is where the recommended range for intra-abdominal and colorectal
pressure ends. Release the key for 2 seconds, then pressure can be increased
above 15 mmHg.
CAUTION!
Exceeding this safety limit is to be decided by and the responsibility of the user/
operator.
5. Start insufflation with the
START
key.
NOTE!
For the safety of the patient, it is advised to start intra-abdominal insufflation
with a Veress needle and at the lowest flow rate (level 1, key
(1)
).
6.
Initial Insufflation with a Veress Needle or Conventional Cannula
Note that the use of a Veress needle is automatically detected by the device. The
AirSeal® recirculation gas flow is NOT ACTIVATED when performing initial insuf-
flation with a Veress needle in the
AirSeal Mode
.The status line depicts
Initial In-
sufflation
. Initial Insufflation ends when the AirSeal® Cannula is attached to the
Tri-Lumen tube set.
Once the set pressure is achieved, (message appears on screen- Connect Tri-Lu-
men Filtered Tube Set to AirSeal® Cannula…)
• Insert the AirSeal® Access Port with inserted obturator.
• Detach the tri-lumen tube from the single lumen adapter.
• Remove the manifold plug.
• Attach the tri-lumen tube to the AirSeal® Access Port bullseye manifold and
tighten.
• The iFS will automatically launch into AirSeal Mode once the tri-lumen tube set
is connected.
• When stabilized, the display will indicate to remove the obturator.
• Upon removal of the obturator, the iFS will perform a calibration. This will take
approximately 8-15 seconds.
• Once complete, an audible tone will sound alerting the user that AirSeal Mode
is active and ready for use and a green bar will appear on the iFS screen show-
ing AirSeal Active.
CAUTION!
If the obturator is removed from the AirSeal® cannula before the desired pres-
sure is achieved, pressure in the cavity may not be maintained.
Содержание AirSeal iFS
Страница 1: ...EN Instructions for use...
Страница 4: ......
Страница 6: ...2 EN 15 AirSeal iFS Warranty 52 16 Test Log 53 16 1 Test Log 53 16 2 Return Form 54 Index 55...
Страница 20: ...16 Initial Device Startup EN House gas supply pressure OK House gas supply pressure too low...
Страница 57: ...Test Log 53 EN 16 Test Log 16 1 Test Log Date Results Comment Signature...
Страница 60: ...W55 186 EN AA...