Initial Device Startup
15
EN
1. Attach the high-pressure hose to the gas connection
2. Fasten the high-pressure hose with the nut
3. Tighten the nut
4. Power on device
5. Select MENU
6. Select Gas supply and choose HOUSE GAS Mode in menu
NOTE!
The default gas settings are selected at the factory and should be changed as
necessary by the user. If a gas supply of greater than 15 bars/218 psi is supplied
to the device, regardless if the default setting is set to bottle gas or house gas,
the device will automatically default to bottle gas mode.
4.2.3
Gas Consumption Display
The gas consumption display indicates the insufflated volume of CO
2
in liters
since the last resetting of the display. The display depicts values between 0 and
999 liters.
The gas consumption display can be reset by pressing the
RESET
key and then re-
turns to
0
.
Gas supply displays
The status of the gas supply is monitored by the device and indicated with sym-
bols and acoustic signals.
Gas supply with gas bottle
The following gas bottle pressures are displayed:
House gas supply
The following house gas supply pressures are displayed:
> 40 bar/580 psi
30 - 40 bar/435 - 580 psi
15 - 30 bar/218-435 psi;
5 - 15 bar/73-218 psi; warning screen "Gas level low. Prepare to
change gas bottle." appears and audible signals can be heard.
If insufflation is stopped, the warning message "Change gas bottle!"
appears and insufflation cannot be restarted until the pressure is >
15 bar/218 psi.
< 5 bar/73 psi; If insufflation is started, the warning message
"Change gas bottle!" appears and acoustic signals (beeps) are emit-
ted. The gas bottle should be changed immediately.
If insufflation is stopped, the warning message "Change gas bottle!"
appears and insufflation cannot be restarted.
Smoke evacuation level will switch to LOW until tank is replaced.
While in AirSeal mode, a countdown of 100 s is displayed during
which the empty gas bottle can be changed while maintaining
abdominal pressure.
0 bar/0 psi; If insufflation is started, the warning message "Change
gas bottle!" Insufflation stopped!" appears and acoustic signals
(beeps) are emitted. The gas bottle should be changed immediately.
If insufflation is stopped, the warning message "Change gas bottle!"
appears and insufflation cannot be restarted.
Содержание AirSeal iFS
Страница 1: ...EN Instructions for use...
Страница 4: ......
Страница 6: ...2 EN 15 AirSeal iFS Warranty 52 16 Test Log 53 16 1 Test Log 53 16 2 Return Form 54 Index 55...
Страница 20: ...16 Initial Device Startup EN House gas supply pressure OK House gas supply pressure too low...
Страница 57: ...Test Log 53 EN 16 Test Log 16 1 Test Log Date Results Comment Signature...
Страница 60: ...W55 186 EN AA...