Informational, Warning and Error Messages
EN
12
Informational, Warning and Error Messages
Status line display
Information for users displayed in the status display are depicted only temporar-
ily and disappear again after a few seconds.
Informational, warning and error messages
Depending on the type of message, windows with green, yellow, and red frames
are depicted.
Informational message
Green frame
Warning message
Yellow frame
Error message
Red frame
Critical message
Red frame
Informational message
Cause
AirSeal is shutting down. System Auto
Calibration in progress.
Pressure is slowly reduced. Pump keeps running for a few seconds. No pressure build
up.
Do not turn device off. Wait for calibration.
Connect blue tube. Smoke evacuation
initiating...
To enable the smoke gas evacuation function, connect the blue tube to a trocar and
open the valve.
Warning message
Cause
Remedy
Check for excessive leakage or suction
Long suction or leaks.
Check for possible leakage! Use suction
carefully.
Fluid Detected in Filter. Change Tube Set
to Protect iFS Unit and Continue with
Smoke Evacuation Function.
Low fluid level of the fluid trap in the
filter housing during Smoke Evacua-
tion mode.
Please insert a new tube set.
Fluid Detected in Filter. Insert Obturator
into AirSeal® Cannula to Maintain Pres-
sure. Change Tube Set to Protect iFS Unit
and Continue with AirSeal Function.
Low fluid level of the fluid trap in the
filter housing during AirSeal mode.
Please insert a new tube set.
Smoke Evacuation stopped. Insufflation
active.
Leaks or low cavity pressure.
Check for possible leakage! Use suction
carefully.
Smoke Evacuation stopped. Suction line
blocked!
Evacuation blocked.
Check the evacuation/suction system for
proper function.
Safety threshold: Pressure >15 mmHg!
The set value of 15 is reached and the
user tries to increase the pressure fur-
ther.
Release the key for 2 seconds. Now you
can set a value up to 20 mmHg.
Starting... Do not install tube set until
self test is completed!
Initial self test is "running".
Wait until the initial self test has com-
pleted.
Occlusion!
Cannula tubing or Veress needle has a
blockage.
Verify cannula / needle placement and
tubing continuity.
Check House Gas Supply!
Delivered Pressure is too low.
Check connection from device to house
gas. Make sure house gas is working prop-
erly.
Connect blue tube. Smoke Evacuation
initiating... Occlusion!
Cannula tubing or Veress needle has a
blockage.
Verify cannula / needle placement and
tubing continuity
Connect Tri-Lumen Filtered Tube Set to
AirSeal® Cannula with Obturator in
Place. AirSeal will Start Automatically.
Occlusion!
Cannula tubing or Veress needle has a
blockage.
Verify Cannula / needle placement and
tubing continuity. Continue with connec-
tion of tubing and cannula. AirSeal will
start automatically.
Содержание AirSeal iFS
Страница 1: ...EN Instructions for use...
Страница 4: ......
Страница 6: ...2 EN 15 AirSeal iFS Warranty 52 16 Test Log 53 16 1 Test Log 53 16 2 Return Form 54 Index 55...
Страница 20: ...16 Initial Device Startup EN House gas supply pressure OK House gas supply pressure too low...
Страница 57: ...Test Log 53 EN 16 Test Log 16 1 Test Log Date Results Comment Signature...
Страница 60: ...W55 186 EN AA...