Operating the Device
19
EN
Press
HOME
key to return to
Mode Selection
.
5.5.1
Selecting Operating Mode
Press the corresponding key in
Mode Selection
to select the desired operating
mode.
The device can be used with three different insufflation operating modes:
1.
AirSeal Mode:
Insufflation with
Tri-Lumen Filtered Tube Set
and AirSeal® Ac-
cess Port.
2.
Smoke Evacuation Mode:
Traditional insufflation and smoke evacuation with
a
Bifurcated Smoke Evac Filtered Tube Set
for two conventional cannulas.
3.
Standard Insufflation Mode:
Traditional insufflation with a
Single Lumen Fil-
tered Tube Set
and conventional cannula.
5.5.2
Insufflation Tube Sets
Please use the indicated accessories when working in the respective modes.
A conventional tube for
Standard Insufflation Mode
can be connected to the
front of the device (see Fig. 5-1 "Front of the Device"
(5)
). The tube sets for the
operating modes
Smoke Evacuation
and
AirSeal
are connected to the filter recep-
tacle (see Fig. 5-1 "Front of the Device"
).
Single Lumen Filtered Tube
Set (hereafter referred to as
single lumen tube)
• Disposable conventional single lumen tube
with standard insufflation filter
Bifurcated Smoke Evac Fil-
tered Tube Set (hereafter
referred to as bifurcated
tube)
Disposable insufflation tube set with filter
• Dual-lumen bifurcated tube, transparent
tube for insufflation, blue tube for smoke
evacuation
Tri-Lumen Filtered Tube Set
(hereafter referred to as tri-
lumen tube)
Disposable insufflation tube set with filter for
use with:
• Tri-lumen tube only for use with AirSeal®
Access Port
SurgiQuest Single Lumen
Adapter
AirSeal® Single Lumen Adapter
• Short section of tube for use with a Veress
needle or conventional cannula with the Tri-
Lumen Filtered Tube Set
SurgiQuest AirSeal® Blade-
less Optical Tip Obturator
(hereafter referred to as
Optical Obturator)
Filter
Single lumen tube
Filter
Bifurcated tube
Filter
Tri-lumen tube
connects to
Single Lumen Adapter
connects to
tri-lumen tube
Veress needle
Содержание AirSeal iFS
Страница 1: ...EN Instructions for use...
Страница 4: ......
Страница 6: ...2 EN 15 AirSeal iFS Warranty 52 16 Test Log 53 16 1 Test Log 53 16 2 Return Form 54 Index 55...
Страница 20: ...16 Initial Device Startup EN House gas supply pressure OK House gas supply pressure too low...
Страница 57: ...Test Log 53 EN 16 Test Log 16 1 Test Log Date Results Comment Signature...
Страница 60: ...W55 186 EN AA...