I
NSTALLATION
.............................
Sievers M500e TOC Analyzer
Operation and Maintenance Manual
DLM 78510-02 EN Rev. B
75
SUEZ © 2022
6.
Locate the terminal block on the back of the rotary power switch inside the
power supply area of the Analyzer.
7.
Connect the AC main wires to the front (the side facing the user) of the
terminal block on the rotary AC power switch, as follows:
•
Connect the
Line
(brown or black wire, typically) wire to the front of the
Terminal block in terminal #5.
•
Connect the
Neutral
(blue or white wire, typically) in terminal #4.
When connecting each wire to the terminal block, loosen the screws (at the
location shown with the arrows) and insert the wires as designated. Tighten
the screws to secure the wires.
8.
After connecting the wires, pull on each connection gently to make sure the
connection is secure.
9.
Attach the AC mains earth ground wire to the grounding lug.
#4
#5