Sievers M500e TOC Analyzer
Operation and Maintenance Manual
DLM 78510-02 EN Rev. B
128
SUEZ © 2022
Running a Grab Sample
Use the
Home
h
screen to perform a grab sample analysis using a vial inserted in a
Standard iOS. (Grab sampling is not available for configurations with the Sample Inlet
Block.)
Alternatively, grab samples can be started from the
User Defined Methods
tab on the
Home
screen.
NOTE:
When measuring in Grab mode and using glass vials, the
sample water must be acidified to a pH of three (3). If you are using
plastic vials, no acidification is required.
To run a grab sample
1. On the
Home
h
screen, verify that a
Grab
method (measurement mode)
appears in the
Protocol
field.
2.
Insert the sampling vial into the vial port.
3.
Click
Start Analysis
.
NOTE:
A grab sample can also be started on the User Defined
Method tab. With the Grab method selected in the Measurement
Mode field, click
Start
. Clicking
Start
is like clicking Start Analysis
on the Home
h
screen. See step
After the Analyzer completes analysis,
Stopped
appears in the
Status
field.