M
AINTENANCE
.............................
Sievers M500e TOC Analyzer
Operation and Maintenance Manual
DLM 78510-02 EN Rev. B
275
SUEZ © 2022
Figure 8-1: Interior Overview of the Analyzer
Table 8-1: Replacement Schedule for Consumables
Description
Typical Operating Life
DI Water Level
6 months
(Also check when replacing other items, and refill as needed)
UV Lamp
6 months
Fan Filter
6 months
Sample Pump Heads
12 months
Resin Cartridge
Typically 24 Months (depending on water quality)
In-line Filter
Replace as Needed (depending on water quality)
Resin
Cartridge
DI Water
Reservoir
Sample
Pump
Heads
UV
Reactor
UV
Lamp
Holder
Fan
Filter