M
AINTENANCE
.............................
Sievers M500e TOC Analyzer
Operation and Maintenance Manual
DLM 78510-02 EN Rev. B
283
SUEZ © 2022
NOTE:
Each port (Prime, Run, and DI Pump Outlet) has a metal latch to
use when attaching or detaching a fitting. Press the latch with a finger to
release a fitting and close the valve. When inserting or detaching the
resin cartridge, insert or detach both end-fittings simultaneously to avoid
damaging the cartridge.
7.
Close and latch the door to the Analyzer.
8.
Re-establish mains power to the Analyzer and turn the power switch to the
ON
position. This will turn on the DI water pump. Let the Analyzer run for 30
minutes to allow the DI pump to circulate water in the DI Loop.
9.
Turn off the Analyzer and open the door to the Analyzer.
10.
Check the level of water in the DI water reservoir. If the water level has
dropped below the
Full
line, remove the rubber inlet cover from the top of the
DI water reservoir, add more DI water, and replace the inlet cover.
11.
Remove the resin cartridge from the
PRIME
position. While holding the resin
cartridge, simultaneously press the latches on the
PRIME
port and
DI Pump
Outlet
port to eject the cartridge.
DI Pump Outlet