M
AINTENANCE
.............................
Sievers M500e TOC Analyzer
Operation and Maintenance Manual
DLM 78510-02 EN Rev. B
288
SUEZ © 2022
Cleaning the Analyzer
When replacing consumables, clean the Analyzer’s external housing and touchscreen as part
of routine preventative-maintenance schedule.
To clean the external housing
Obtain a clean, soft cloth dampened with water or a non-abrasive cleaner, and another
clean
dry
soft cloth.
1. If the Analyzer is taking measurements, click the
Stop Analysis
button.
2.
On the Analyzer’s Home screen, click the
Power Off
button to close all the
files.
3.
Turn the rotary power switch (located on the left outside panel of the Analyzer)
to the
Off
position to complete the power off cycle.
4.
Turn off power to the Analyzer with the main power switch.
5.
Wipe the external housing with a clean cloth dampened with water or a non-
abrasive cleaner.
6.
Wipe dry with another clean and
dry
soft cloth.
To clean the touchscreen
1. Obtain a clean, soft (and
dry
) cloth or cotton pad.
2.
Navigate to the
System Tests
panel (
Settings
screen >
Diagnostics
tab).
3.
On the scrolling menu, select
Clean Screen
and click
Run
to temporarily
disable the screen for 30 seconds. Repeat, as needed, if the 30-second time
frame expires while cleaning.
4.
Wipe the touchscreen with a clean, soft (and dry) cloth or cotton pad.
Table 8-2: Sievers M500e TOC Analyzer 3
-
Year Maintenance Worksheet
Item
Period
Installation Date
Notes
UV Lamp
6 months
DI Water Level*
6 months
Fan Filter
6 months
Pump Heads
12 months
UV Lamp
12 months
DI Water Level*
12 months