Sievers M500e TOC Analyzer
Operation and Maintenance Manual
DLM 78510-02 EN Rev. B
153
SUEZ © 2022
1. On the
Settings
s
screen, select the
Hardware
tab.
2.
On the scrolling menu, select
Alarms Setup
and click
Run
. The
Alarms Setup
screen appears.
3.
In the
Select Alarm
area, select one of the following options to configure
(
Alarm
1
,
Alarm 2
,
Alarm 3
,
or
Alarm 4
).
4.
Click the
Threshold
field and use the numeric keypad to set a value for the
alarm threshold (1-3000 ppb) and click
OK
.
5.
In the table area, select an assignment option for applying the alarm (such as
TOC>Threshold
,
Error or Warning
, or
Temperature > Threshold
.
6.
Repeat steps
7.
Click
Save
.