I
NSTALLING
THE
RESIN
CARTRIDGE
.............................
Sievers M500e TOC Analyzer
Operation and Maintenance Manual
DLM 78510-02 EN Rev. B
312
SUEZ © 2022
3.
Turn the power switch to the
O
N
position. This will turn on the DI water pump.
Let the Analyzer run 180 minutes to allow the DI pump to circulate water
through the DI Loop.
4.
After 180 minutes, on the
Home
screen, click the
Power On/Off
button.
5.
Move the resin cartridge from the
Prime
position to the
Run
position. Press
the top latch of the resin cartridge to release it and gently rotate the top of
cartridge to the
Run
position. Push to insert the elbow-fitting into the
Run
port,
while leaving the lower elbow-fitting end in the DI pump outlet.
Verify that the ends are secure before moving to the next step.
6.
View the water level in the DI reservoir and verify that the water level reaches
the FULL line, as indicated on the outside of the reservoir.
The water level may now be lower, as water circulates in the DI Loop and
saturates the resin cartridge during the priming process. If needed, fill the DI
reservoir with additional low-TOC DI water to reach the FULL line, indicated
on the outside of the reservoir. There is no need to prime the DI pump again.
7.
Close and latch the Analyzer’s door.
DI Pump Outlet