Accuracy & Precision Verification Protocol
1.0 I
NTRODUCTION
Perform the following Accuracy & Precision Protocol to verify the TOC accuracy and
precision of a Sievers M500e TOC Analyzer.
2.0 S
COPE
This procedure applies to the Sievers M500e TOC Analyzer.
3.0 M
ATERIALS
NOTE:
Use
only
Sievers standards purchased from SUEZ.
3.1 Sievers M500e TOC Analyzer
3.2
Accuracy & Precision Verification Worksheet
3.3
Accuracy/Precision Verification Standard Set
:
3.3.1 One vial —
Rw Blank
3.3.2 One vial —
500 ppb TOC
4.0 D
EFINITIONS
4.1 None
5.0 P
ROCEDURE
5.1 If the Analyzer is taking measurements, click
Stop Analysis
.
5.2 On the
System Protocols
p
screen, select the
Verification
tab.