I
NSTALLATION
.............................
Sievers M500e TOC Analyzer
Operation and Maintenance Manual
DLM 78510-02 EN Rev. B
63
SUEZ © 2022
S
TEP
4: I
NSTALL
I/O C
ONTROL
W
IRING
(O
PTIONAL
)
The Sievers M500e TOC Analyzer offers three 4-20 mA outputs, four alarm outputs, and
two isolated binary inputs via the four terminal blocks on the instrument controller board
(located inside and at the top of the Analyzer). The following terminal block positions are
“To install I/O wiring” on page 64
•
J1 — Alarm Relay Outputs
•
J2 — Isolated State Inputs
•
J3 — Isolated State Outputs (This terminal is not used. Do
not
make any
connections to this terminal block.)
•
J4 — 4-20 mA Analog Current Loop
This section includes step-by-step instructions for
“Installing I/O Wiring” on page 63
.
Additional information to assist in the installation is also provided:
•
“Working with Alarm Outputs (J1)” on page 68
•
“Working with 4-20 mA Analog (J4) Outputs” on page 69
•
“Working with Binary Inputs (Remote Start/Stop) (J2)” on page 71
After installing I/O control wiring, configure the analog and alarm settings, and activate
the binary input, in the Analyzer’s firmware later in
" Step 9: Configure Basic Analyzer
Installing I/O Wiring
Use this instruction to install the following on the I/O instrument controller board, as
needed:
•
4-20 mA output connection
•
Alarm output connections
•
Binary (start/stop) input connections