82
SSCI - Suburban Surgical Company, Inc.
Chapter 5 - Repairs & Replacements
Seal, IV Port
P/N 752100
There are four IV ports on the Intensive Care Unit; two over each
half-size compartment. Each IV port consists of a seal, a seal
retainer, two retaining screws, and two locknuts. All four IV port
seals are identical and are removed and installed the same way.
Tools Required
11/32-in. wrench
Phillips screwdriver
Removal
1.
Open both animal compartment doors.
2.
Use a 11/32-in. wrench and a Phillips screwdriver to
remove the two cap screws that hold the seal retainer in
place (Figure 91) and remove the retainer from the unit.
3.
Remove the IV Seal from the unit.
Installation
Note:
On the seal and seal retainer, the mounting holes are closer
to the top edge than to the bottom edge (Figure 92).
1.
Place the IV seal behind the IV port opening with the short
side on top.
2.
Hold the seal retainer in place, flanges outward, over the
seal and secure with the two cap screws and locknuts.
3.
Close the doors.
Figure 91. Removing the IV Port
Cap Screws
Figure 92. Parts of the IV Port
Seal
Cap
Screws
Nuts
Seal
Retainer
Short
Long
Top
Содержание REGAL INTENSIVE CARE UNIT 12019-01-NGFREP
Страница 12: ...x SSCI Suburban Surgical Company Inc Table of Contents Notes...
Страница 20: ...8 SSCI Suburban Surgical Company Inc Chapter 1 General Information Notes...
Страница 44: ...32 SSCI Suburban Surgical Company Inc Chapter 2 Description of Components Notes...
Страница 148: ...136 SSCI Suburban Surgical Company Inc Chapter 5 Repairs Replacements Notes...
Страница 168: ...156 SSCI Suburban Surgical Company Inc Chapter 6 Troubleshooting Notes...
Страница 170: ...A 2 SSCI Suburban Surgical Company Inc Appendix Thispageleftblankintentionally...
Страница 174: ...A 6 SSCI Suburban Surgical Company Inc Appendix Thispageleftblankintentionally...
Страница 183: ...Inside back cover This page left blank intentionally...