Form No. 702718 Rev. A March, 2005
127
Owner’s Manual - Regal Intensive Care Unit
Digital Readout
P/N 854591
Tool Required
Phillips screwdriver
Removal
1.
Open the filter/fans compartment (refer to
Page 58
).
2.
With a Phillips screwdriver, un-screw the two screws that
hold the digital readout mounting bracket to the rear of the
panel (Figure 148).
3.
With a Phillips screwdriver, remove the two screws that
hold the readout to the bracket.
Installation
1.
Mount the bracket to the new readout with the two screws
removed above (Figure 148).
2.
Mount the readout/bracket assembly to the panel with the
two screws removed above.
3.
Place the
F° - C°
switch to the desired setting.
4.
Make sure there is a battery in the readout. If not, refer to
Page 62.
5.
Replace and secure the filter/fans compartment panel.
Figure 148. Digital Readout and Mounting Bracket
Digital
Readout
Mounting
Bracket
Mounting
Bracket-Panel
Screw (2)
Readout -
Mounting
Bracket
Screw (2)
F° - C°
Switch
Содержание REGAL INTENSIVE CARE UNIT 12019-01-NGFREP
Страница 12: ...x SSCI Suburban Surgical Company Inc Table of Contents Notes...
Страница 20: ...8 SSCI Suburban Surgical Company Inc Chapter 1 General Information Notes...
Страница 44: ...32 SSCI Suburban Surgical Company Inc Chapter 2 Description of Components Notes...
Страница 148: ...136 SSCI Suburban Surgical Company Inc Chapter 5 Repairs Replacements Notes...
Страница 168: ...156 SSCI Suburban Surgical Company Inc Chapter 6 Troubleshooting Notes...