Chapter 6 - Troubleshooting
Cannot move the set-point needle in the oxygen
concentration meter.
CAUTION:
Before working on the electrical system,
make sure the unit is Off and the electric power cord
un-plugged.
First:
A wire connection from the potentiometer to the oxygen
concentration meter may have come loose from the meter. Refer to
Figure 101 on
Page 93
and make sure that both connections are in
place and tight.
Second:
A wire on the rear of the potentiometer may be broken
off (Figure 110 on
Page 97
). Have a person skilled in soldering re-
solder the wire in place. If this is not possible, replace the
potentiometer. Order P/N 853397 and refer to
Page 93
for
replacement instructions.
Third:
The potentiometer itself maybe be bad. Replace the
potentiometer. Order P/N 853397 and refer to
Page 96
for
replacement instructions.
Remedial Action
Содержание REGAL INTENSIVE CARE UNIT 12019-01-NGFREP
Страница 12: ...x SSCI Suburban Surgical Company Inc Table of Contents Notes...
Страница 20: ...8 SSCI Suburban Surgical Company Inc Chapter 1 General Information Notes...
Страница 44: ...32 SSCI Suburban Surgical Company Inc Chapter 2 Description of Components Notes...
Страница 148: ...136 SSCI Suburban Surgical Company Inc Chapter 5 Repairs Replacements Notes...
Страница 168: ...156 SSCI Suburban Surgical Company Inc Chapter 6 Troubleshooting Notes...
Страница 170: ...A 2 SSCI Suburban Surgical Company Inc Appendix Thispageleftblankintentionally...
Страница 174: ...A 6 SSCI Suburban Surgical Company Inc Appendix Thispageleftblankintentionally...
Страница 183: ...Inside back cover This page left blank intentionally...