104
SSCI - Suburban Surgical Company, Inc.
Chapter 5 - Repairs & Replacements
System Light
P/N 854589
There are six red system lights on the electronics compartment
door; one each for alarm, oxygen, cooling, heat, left floor tempera-
ture, and light floor temperature. All lights are identical and are
removed and installed the same way.
CAUTION:
Before proceeding, make sure the unit is
Off, and the electric power cord un-plugged.
Removal
1.
Make sure that the main on/off switch is Off and the
amber light in the switch is out.
2.
Unplug the power cord from the electrical outlet.
3.
Open the electronics compartment (refer
to
Page 57
).
4.
Pull both wires off of the rear of the light
(Figure 116).
5.
Squeeze all four “ears” inward, and push
the light out through the front of the door.
Installation
1.
Push the light body into the door from the
front until the “ears” snap into place.
2.
Re-place both wires onto the light terminals.
Either wire can go to either terminal.
3.
Close and secure the electronics compartment door.
4.
Plug the power cord into its wall outlet.
Figure 116. Inside View of System Light
“Ear” (1 of 4)
Wire
Connections
Содержание REGAL INTENSIVE CARE UNIT 12019-01-NGFREP
Страница 12: ...x SSCI Suburban Surgical Company Inc Table of Contents Notes...
Страница 20: ...8 SSCI Suburban Surgical Company Inc Chapter 1 General Information Notes...
Страница 44: ...32 SSCI Suburban Surgical Company Inc Chapter 2 Description of Components Notes...
Страница 148: ...136 SSCI Suburban Surgical Company Inc Chapter 5 Repairs Replacements Notes...
Страница 168: ...156 SSCI Suburban Surgical Company Inc Chapter 6 Troubleshooting Notes...
Страница 170: ...A 2 SSCI Suburban Surgical Company Inc Appendix Thispageleftblankintentionally...
Страница 174: ...A 6 SSCI Suburban Surgical Company Inc Appendix Thispageleftblankintentionally...
Страница 183: ...Inside back cover This page left blank intentionally...