SSCI - Suburban Surgical Company, Inc.
Chapter 4 - Operation and Care
Replacing the
Fuel Cell
CAUTION:
Do not remove the fuel cell from the bag
until ready to use.
CAUTION:
Wear safety glasses when handling the
fuel cell.
CAUTION:
Avoid contact with the sensing surface.
CAUTION:
The fuel cell contains lead, a chemical
known to cause cancer, birth defects, or other
reproductive harm.
CAUTION:
Wash your hands thoroughly after
handling the fuel cell.
The fuel cell powers the black calibration needle in the oxygen
concentration meter. The cell has an average life of about one year,
depending on frequency of use. Difficulty in moving the black
calibration needle in the oxygen concentration meter is usually an
indication of an exhausted fuel cell. Order replacements from SSCI,
P/N 853396. To replace the fuel cell:
1.
Open the brown box, then open the plastic bag inside and
take out the fuel cell.
2.
Open the electronics compartment (refer to
Page 57
).
3.
Open the filter/fans compartment (refer to
Page 58
).
4.
Inside the filter/fans compartment, notice the installation of
the oxygen sensor (Figure 79).
Figure 78. Oxygen Sensor Inside the
Electronics Compartment
Oxygen Sensor
Spiral
Cord
Figure 77. Cooling Coil in Rear of Unit
Содержание REGAL INTENSIVE CARE UNIT 12019-01-NGFREP
Страница 12: ...x SSCI Suburban Surgical Company Inc Table of Contents Notes...
Страница 20: ...8 SSCI Suburban Surgical Company Inc Chapter 1 General Information Notes...
Страница 44: ...32 SSCI Suburban Surgical Company Inc Chapter 2 Description of Components Notes...
Страница 148: ...136 SSCI Suburban Surgical Company Inc Chapter 5 Repairs Replacements Notes...
Страница 168: ...156 SSCI Suburban Surgical Company Inc Chapter 6 Troubleshooting Notes...
Страница 170: ...A 2 SSCI Suburban Surgical Company Inc Appendix Thispageleftblankintentionally...
Страница 174: ...A 6 SSCI Suburban Surgical Company Inc Appendix Thispageleftblankintentionally...
Страница 183: ...Inside back cover This page left blank intentionally...