106
SSCI - Suburban Surgical Company, Inc.
Chapter 5 - Repairs & Replacements
6.
Remove the printed circuit board/display assembly from the
rear of the unit.
7.
Pull the bezel out from the front of the door (Figure 118).
Installation
1.
Place the threaded studs on the new bezel into the four
screw holes on the front of the door.
2.
Inside the door, place the printed circuit board/display
assembly over the four studs on the bezel. The ex-
tended tab portion of the circuit board should point down.
If you install it with the tab pointing up, the display will be
upside-down!
3.
Secure the assembly together with the four nuts removed
above.
Note:
Four spare nuts are provided with the new
display. None of the other hardware items supplied in the
bag assembly will be needed for this installation.
4.
Observe the ribbon cable on the 4-wire connector removed in
Removal - Step 4
above. Note that one wire has red stripes.
5.
Connect the 4-wire ribbon connector to the rear of the
display. The wire with the red stripe should be to your right.
Note:
A new ribbon cable is supplied with the display.
You can use the existing cable, or install the new one.
6.
Close and secure the electronics compartment door.
7.
Plug the power cord into its wall outlet.
Содержание REGAL INTENSIVE CARE UNIT 12019-01-NGFREP
Страница 12: ...x SSCI Suburban Surgical Company Inc Table of Contents Notes...
Страница 20: ...8 SSCI Suburban Surgical Company Inc Chapter 1 General Information Notes...
Страница 44: ...32 SSCI Suburban Surgical Company Inc Chapter 2 Description of Components Notes...
Страница 148: ...136 SSCI Suburban Surgical Company Inc Chapter 5 Repairs Replacements Notes...
Страница 168: ...156 SSCI Suburban Surgical Company Inc Chapter 6 Troubleshooting Notes...
Страница 170: ...A 2 SSCI Suburban Surgical Company Inc Appendix Thispageleftblankintentionally...
Страница 174: ...A 6 SSCI Suburban Surgical Company Inc Appendix Thispageleftblankintentionally...
Страница 183: ...Inside back cover This page left blank intentionally...