Form No. 702718 Rev. A March, 2005
35
Owner’s Manual - Regal Intensive Care Unit
Installing the
Fuel Cell
A fuel cell in the oxygen sensor powers the black calibration needle
in the oxygen concentration meter. The fuel cell comes packed in a
brown, cardboard box marked “Teledyne Oxygen Sensor”
(Figure 35).
CAUTION:
Do not remove the fuel cell from the bag
until ready to use.
CAUTION:
Wear safety glasses when handling the
fuel cell.
CAUTION:
Avoid contact with the sensing surface.
CAUTION:
The fuel cell contains lead, a chemical
known to cause cancer, birth defects, or other
reproductive harm.
CAUTION:
Wash your hands thoroughly after
handling the fuel cell.
1.
Open the brown box, then open the plastic bag inside and
take out the fuel cell (Figure 35).
2.
Open the electronics compartment (refer to
Page 57
).
3.
Open the filter/fans compartment (refer to
Page 58
).
4.
Inside the filter/fans compartment, notice the installation of
the oxygen sensor (Figure 37).
Figure 35. Fuel Cell Box and Bagged
Fuel Cell
Figure 36. Oxygen Sensor Inside the
Electronics Compartment
Oxygen Sensor
Spiral
Cord
Содержание REGAL INTENSIVE CARE UNIT 12019-01-NGFREP
Страница 12: ...x SSCI Suburban Surgical Company Inc Table of Contents Notes...
Страница 20: ...8 SSCI Suburban Surgical Company Inc Chapter 1 General Information Notes...
Страница 44: ...32 SSCI Suburban Surgical Company Inc Chapter 2 Description of Components Notes...
Страница 148: ...136 SSCI Suburban Surgical Company Inc Chapter 5 Repairs Replacements Notes...
Страница 168: ...156 SSCI Suburban Surgical Company Inc Chapter 6 Troubleshooting Notes...
Страница 170: ...A 2 SSCI Suburban Surgical Company Inc Appendix Thispageleftblankintentionally...
Страница 174: ...A 6 SSCI Suburban Surgical Company Inc Appendix Thispageleftblankintentionally...
Страница 183: ...Inside back cover This page left blank intentionally...