Chapter 6 - Troubleshooting
The alarm system does not work.
First:
Make sure that the main on/off switch and the oxygen on/
off switch are On. Both these switches must be On for the alarm
system to operate.
Second:
Make sure that the alarm time delay relay is seated
securely in place and is properly set. Refer to
Setting the Alarm
Time Delay Relay
on
Page 59.
Also, make sure the four wires are
securely connected to the octal base under the relay.
Third:
When the alarm on/off switch is On, the amber light in the
switch should be on. If this light does not illuminate, check the wire
connections on the back of the switch. If they are all secure, you
may have a bad switch. Order P/N 853422 and refer to
Page 99
for replacement instructions.
Fourth:
Check the wire connections at the rear of the audible
alarm. If they are secure, the audible alarm may be defective. Order
P/N 853458 and refer to
Page 116
for replacement instructions.
Remedial Action
Содержание REGAL INTENSIVE CARE UNIT 12019-01-NGFREP
Страница 12: ...x SSCI Suburban Surgical Company Inc Table of Contents Notes...
Страница 20: ...8 SSCI Suburban Surgical Company Inc Chapter 1 General Information Notes...
Страница 44: ...32 SSCI Suburban Surgical Company Inc Chapter 2 Description of Components Notes...
Страница 148: ...136 SSCI Suburban Surgical Company Inc Chapter 5 Repairs Replacements Notes...
Страница 168: ...156 SSCI Suburban Surgical Company Inc Chapter 6 Troubleshooting Notes...
Страница 170: ...A 2 SSCI Suburban Surgical Company Inc Appendix Thispageleftblankintentionally...
Страница 174: ...A 6 SSCI Suburban Surgical Company Inc Appendix Thispageleftblankintentionally...
Страница 183: ...Inside back cover This page left blank intentionally...