Form No. 702718 Rev. A March, 2005
151
Owner’s Manual - Regal Intensive Care Unit
Excessive carbon dioxide in the animal compartment.
First:
The Sodalime supply in the filter/fans compartment is
probably exhausted. Replace the Sodalime with fresh material.
Refer to
Refilling the CO
2
Filter
on
Page 65.
Second:
Make sure the circulating fans in the filter/fans
compartment are working. Refer to
Circulating fan does not work
on
Page 149.
Third:
Make sure that the air inlet grill and the air outlet grill
(Figure 3 on
Page 9
) are clear and unobstructed.
Fourth:
Make sure that your set-point setting on the oxygen
concentration meter is correct. Refer to
Using Oxygen
on
Page 51
.
Fifth:
Make sure that the flow rate on the oxygen flow meter is
correct. Refer to
Controlling Oxygen Flow Rate
on
Page 56
.
Sixth:
Check your oxygen supply and pressure regulator setting.
Seventh:
Refer to
Oxygen flow is inadequate
on
Page 141.
Remedial Action
Содержание REGAL INTENSIVE CARE UNIT 12019-01-NGFREP
Страница 12: ...x SSCI Suburban Surgical Company Inc Table of Contents Notes...
Страница 20: ...8 SSCI Suburban Surgical Company Inc Chapter 1 General Information Notes...
Страница 44: ...32 SSCI Suburban Surgical Company Inc Chapter 2 Description of Components Notes...
Страница 148: ...136 SSCI Suburban Surgical Company Inc Chapter 5 Repairs Replacements Notes...
Страница 168: ...156 SSCI Suburban Surgical Company Inc Chapter 6 Troubleshooting Notes...
Страница 170: ...A 2 SSCI Suburban Surgical Company Inc Appendix Thispageleftblankintentionally...
Страница 174: ...A 6 SSCI Suburban Surgical Company Inc Appendix Thispageleftblankintentionally...
Страница 183: ...Inside back cover This page left blank intentionally...