Owner’s Manual - Regal Intensive Care Unit
Calibration knob will not move the calibration needle
This is almost always the symptom of an exhausted fuel cell in the
oxygen sensor. Replace the fuel cell. Order P/N 853396 and refer
to
Page 85
for replacement instructions.
Remedial Action