Torq-Tronics 2
®
/
Owners Manual / Appendix A: Serial Commands
A.6.1.14 Show the user string for the tester. Up to 16 characters long type: CMP<Enter>Or
CM#<Enter> These commands are interchangeable. However if using an ExaWin program,
use CM#<Enter>
Example Returns: Tester1234567890<CR+LF>
A.6.1.15 Show the user string for the transducer. Up to 16 characters long type: CMW<Enter>
Example Returns: Test Station 1<CR+LF>
A.6.1.16 Print the model string Type: E<Enter>
Example Returned: TorqTronics II 10Ft.Lb.<CR+LF>
A.6.1.17 Show the auto clear time in seconds Type: H9<Enter>
Example Returns: 0002<CR+LF> for two seconds. (Hexadecimal format: 0 to ‘FF”)
A.6.1.18 Set the auto clear time in seconds. (Hexadecimal format: 0 to ‘FF’).) type: H9xx<Enter>
Example Returns: 0002<CR+LF> for two seconds.
A.6.1.19 Request current calibration table information type: L<Enter>
Returns example:
Table 08 (table location in tester memory)
0040,0025
(Peak drop setting, second value is not used.)
TorqTronics II 600Ft.lb,,1234 (Tester model and serial number.)
CW =A97A,A554,A129,9CF8,98C8,949C,9068,8C3D,880E,83D9,7FA7, (Tester Cw
ADC calibration points 100% to 0%.)
1.07673,1.07570,1.07505,1.07507,1.07492,1.07392,1.07442,1.07226,1.06994,1.06895,
Ccw=55C0,59ED,5E20,6246,6671,6AA5,6ED6,7305,7735,7B66,7F9E,
(Tester Ccw ADC calibration points 100% to 0%.)
1.07573,1.07537,1.07568,1.07421,1.07309,1.07332,1.07292,1.07160,1.06944,1.06402,
Last Cal: 10082012 (Tester last calibration date.)
Next Cal: 10082013 (Tester next suggested calibration date.)
A.6.1.20 Show the mode of operation type: N<Enter>
Returned Examples:
PEAK<CR+LF>
peak mode message.
INIPEAK<CR+LF>
initial peak mode message.
TRACK<CR+LF>
Track mode message.
PTOOL*<CR+LF>
Power tool message.
|__power tool filter in use.
Note: the “*” indicates no filter selected, valid filter selection range: *, 0-9.
A.6.1.21 Set the mode of operation type: Px<Enter>
Examples:
P1<Enter> = TRACK<CR+LF> Track mode message.
P2<Enter> = PEAK<CR+LF> peak mode message.
P3<Enter> = INIPEAK<CR+LF> initial peak mode message.
P4<Enter> = PTOOL*<CR+LF> Power tool message.
|__power tool filter in use.
Note: the “*” indicates no filter selected, valid filter selection range: *, 0-9.
A.6.1.22 Print the current torque reading. String construction controlled by the last state of the TX
command.
Type: P<Enter>