Torq-Tronics 2
® /
Owners Manual
16.4.11 Make certain the calibration arm is properly counterbalanced.
16.4.12 Attach the drive square of the calibration arm to the drive of the tester in the clockwise
direction, and press the Zero key.
16.4.13 Hang the weight platform (hook) and the CW full-scale weights (100%) (Found in the
“Calibration Table”) on the calibration arm-gently, and imparting as little motion to
the weights as possible. Wait until the weights have completely stopped moving then
remove the full-scale load.
16.4.14 Wait one minute.
16.4.15 Repeat steps 12 and 13 above two more times, then proceed to next step.
16.4.16 Remove the calibration arm from the unit under test.
16.4.17 Press the ‘CLEAR’ or ‘DOWN’ key on the unit to capture the CW 0 calibration point.
(Green led will flicker.)
16.4.18. Press the ‘MENU’ or ‘UP’ key to move to the next test point.
16.4.19 Again hang the weight platform (hook) and the appropriate weights (10%) (found in
the “Calibration Chart”) on the calibration arm-gently, and imparting as little motion
to the weights as possible. Wait for the arm to completely stop moving then press the
‘CLEAR’ or ‘DOWN’
key on the unit.
Observe acknowledgement of the 10% clockwise calibration point.
16.4.20 Press the ‘MENU’ or ‘UP’ key to move to the next test point.
16.4.21 Again hang the weight platform (hook) and the appropriate weights (20%) (found in
the “Calibration Chart”) on the calibration arm-gently, and imparting as little motion
to the weights as possible. Wait for the arm to completely stop moving then press the
‘CLEAR’ or ‘DOWN’
key on the unit.
Observe acknowledgement of the 20% clockwise calibration point.
16.4.22 Press the ‘MENU’ or ‘UP’ key to move to the next test point.
16.4.23 Again hang the weight platform (hook) and the appropriate weights (30%) (found in
the “Calibration Chart”) on the calibration arm-gently, and imparting as little motion
to the weights as possible. Wait for the arm to completely stop moving then press the
‘CLEAR’ or ‘DOWN’
key on the unit.
Observe acknowledgement of the 30% clockwise calibration point.
16.4.24 Press the ‘MENU’ or ‘UP’ key to move to the next test point.
16.4.25 Again hang the weight platform (hook) and the appropriate weights (40%) (found in
the “Calibration Chart”) on the calibration arm-gently, and imparting as little motion
to the weights as possible. Wait for the arm to completely stop moving then press the
‘CLEAR’ or ‘DOWN’
key on the unit.
Observe acknowledgement of the 40% clockwise calibration point.
16.4.26 Press the ‘MENU’ or ‘UP’ key to move to the next test point.
16.4.27 Again hang the weight platform (hook) and the appropriate weights (50%) (found in the
“Calibration Chart”) on the calibration arm-gently, and imparting as little motion to the
weights as possible. Wait for the arm to completely stop moving then press the ‘CLEAR’ or
‘DOWN’
key on the unit.
Observe acknowledgement of the 50% clockwise calibration point.
Owners Manual
Torq-Tronics 2
Sturtevant Richmont
Direct: 1 847/455-8677
U.S. Toll-Free: 800/877-1347
www.srtorque.com
11/16/201
7
REV
E
. Document Page 23 of 39