Torq-Tronics 2
® /
Owners Manual
13.11 For more information on understanding and resolving the over capacity situation, see
our videos:
Interview With Torq-Tronics 2
https://www.youtube.com/ watch?v=_8TH3frmKH4
and
Error Proofing With An LED
https:// www.youtube.com/watch? v=QW9068Ewo0k
PLEASE NOTE:
Your Quality Is Our Business is more than an advertising tagline with us, it is our way of life. We take
your quality seriously. Like all Sturtevant Richmont products Torq-Tronics 2 is engineered and built
to withstand the rigors and realities of the industrial workplace.
Our testers were tested at 100% of capacity at more than 50,000 cycles in both directions and the units
stay in spec. However, at 120% of capacity we can no longer be 100% certain that the unit will function
properly.
To safeguard your test results we suggest that you have any Torq-Tronics 2 unit that has been placed
in overload, tested. To test the Torq-Tronics 2 follow the procedures outlined in Section 15: Calibra-
tion, or email us at [email protected] to schedule returning the Torq-Tronics 2 so we can
test it. If the tests prove to be accurate, put the unit back in service and have it calibrated at the next
cycle. The last and next calibration dates are displayed on the display during the start up process.
If your tests reveal the unit may have been compromised, please return it to Sturtevant Richmont for
any required repair, recalibration, and recertification.
14:
14.1 Always use NiMH rechargeable batteries and keep
the polarity correct.
14.2
NEVER mix battery brands, types, ages, as this can damage
the unit.
14.3
When cleaning the unit, NEVER use a spray bottle to apply
cleaning solutions. Extra cleaning solution may drain down
between the keys into the unit and cause electronic damage.
14.4
Always use a soft, damp cloth or sponge to wipe the unit. Do
not use any abrasive cleaners or abrasive surfaces on the unit.
14.5
Always test or calibrate tools within 10% to 75% of Torq-
Tronics 2 capacity. Capacity is located on a plate on the right hand side of the unit. Unit
Capacity is also displayed on the display during the start up sequence.
14.6
The unit holds 999 records in memory. When it exceeds 999 records the unit will over
write previous records. Each record is date and time stamped.
14.7
The Torq-Tronics 2 has safeguards designed to notify the user of overload at 120% of
capacity. It is recommended that the unit be checked for accuracy after an overload
occurrence.
Owners Manual
Torq-Tronics 2
Sturtevant Richmont
Direct: 1 847/455-8677
U.S. Toll-Free: 800/877-1347
www.srtorque.com
Caution: spraying cleaner on
the face of Torq-Tronics 2
may cause electrical damage
to the unit and will void the
warranty.
Please use a soft damp cloth
to clean Torq-Tronics 2.
11/16/201
7
REV
E
. Document Page 18 of 39