Torq-Tronics 2
®
/
Owners Manual / Appendix A: Serial Commands
A.1
Default Factory Communication Settings
A.1.1 Serial communication parameters at boot up: 4800 baud, 8 data bits and 1 stop bit, Parity:
None.
A.2
Clearing ‘Overload’ of Torq-Tronics 2 process
A.2.1 Using a Hyper-terminal type program set for 4800 baud, 8data bits and 1 stop bit,
Parity: None.
A.2.1.1 Type: D<enter>
A.2.1.2 thenType: Z <enter>
A.2.1.3 then Type: D<enter>
A.
3
To See the Captured Overload Point of Torq-Tronics 2 process
A.3.1 Using a Hyper-terminal type program set for 4800 baud, 8data bits and 1 stop bit, Parity:
None.
A.3.1.1 Type: O<enter>
A.3.1.2 The Torq-Tronics2 will send, via the RS-232, a print string showing the captured
overload.
Print string example: (Format controlled by TX command below.)
+18.5649,In.oz,PEAK,NA,005,00001234,Test Station 1,.000001,.000001,05/31/2011,11:57:42<CR
+LF>
| | | | |
|
| | |
| |
|
|_end
of
record.
| | | | |
|
| | |
| |
|_Controlled
by
PDate-
Time & PUDFormat.
| | | | |
|
| | |
| |_Controlled
by
PMax.
| | | | |
|
| | |
|_Controlled
by
PMin.
| | | | |
|
| | |_controlled
by
PUserString.
| | | | |
|
| |_controlled
by
PSerial
Num
| | | | |
|
|_Controlled
by
PFieldSep.
| | | | |
|_Controlled
by
PNumRec
| | | | |_Controlled
by
POkNg
| | | |_controlled
bt
PMode
|
|
|_Controlled by PUnit
|
|_Controlled by PActTor
|_Controlled by PSine.
Owners Manual
Torq-Tronics 2
Sturtevant Richmont
Direct: 1 847/455-8677
Toll-Free: 800/877-1347
www.srtorque.com
Print string example: (Format controlled by TX command below.)
+18.5649,In.oz,PEAK,NA,005,00001234,Test Station 1,.000001,.000001,05/31/2011,11:57:42<CR+LF>
Controlled by PSine.
Controlled by P
ActT
or
Controlled by PUnit
Controlled by PMode Controlled by PNumRec
Controlled by PFieldSep
Controlled by POkNg
Controlled by PUserString
Controlled by PMin.
Controlled by PMax.
Controlled by PDateT
ime
& PUDFormat
Controlled by PSerial Num
end of message