5.3.7.2.1 The ‘DOWN’ arrow key pages through the ten filters available, numbered 0 through
9. Each filter is tuned for a specific frequency of measurement see owner’s manual
page 13, section 11, for further information.
5.3.7.3 Press ‘Peak Torque’ key to save mode setting.
5.3.8
TEST POINT:
used to enter/exit into target value setting menu.
5.3.8.1
Using the ‘UP’ and ‘DOWN’ arrow keys, the operator can set the target torque value
for the test.
5.3.8.1.1 Note: then ‘UNIT’ key is used to change the position of the decimal point
5.3.8.1.2 Note: The ‘ZERO’ key returns the test point to zero.
5.3.8.2
Pressing the ‘TEST POINT’ button a second time saves the value to memory and
enters into the % of tolerance window.
5.3.8.2.1 Be sure to use the appropriate % of tolerance for the tool to be tested.
5.3.9
Using the ‘UP’ and ‘DOWN’ arrow keys.
5.3.9.1
Note: The ‘ZERO’ key returns the test % to 0.1%.
5.3.9.2
Pressing the ‘Test Point’ button a third time saves the settings to memory and exits the
target setting menu.
6:
Operation
Please follow the procedure for each wrench tested. Be sure the Torq-Tronics 2 tester has
been exercised three times to capacity in the direction CW or CCW that it will be used.
Adjustable wrenches should also be exercised three times at 4 different points along the
scale of capacity before beginning the test. Preset wrenches should be tested at preset
value.
Preparation for Testing
6.1
Move the 3 position Power Switch from Off to On or to Battery as needed.
6.2
Allow the tester to complete the internal start up process. This takes about 20 seconds.
6.2.1 At the end of startup process, the tester will return to its last state of operation.
6.2.1.1 Unit of Measure selected.
6.2.1.2 Mode of operation selected. (and filter is applicable.)
6.2.1.3 Torque specifications.
6.3
Auto store/print/clear process is always active except in ‘Track’ mode. When the ‘Clear’
button is pressed the torque data is transmitted via the RS-323 link then cleared from
the display. While in ‘Track’ mode no data is stored in the tester.
6.4
If the torque value Display reads other than zero, press the ‘ZERO’ key.
6.4.1
Note: Never press the Zero key with a wrench connected to the tester. Doing so will
cause inaccurate results. The ‘ZERO’ key does not perform a TARE function.
6.5
6.6
The Torq-Tronics 2 should be used between 10% and
100
% of the unit capacity. Ex-
ample: If the tester is rated at 100 ft. lbs. you should be testing tools with a capacity
between 10 ft lbs and
100
ft. lbs. If you test below 10% of rated capacity you
may
encounter problems with accuracy, especially if you are testing a torque screwdriver
with an accuracy of +/- 6%.
See our video on Set Up For Testing on the Newton Metre Channel in YouTube:
https://www.youtube.com/watch?v=BOLzJ3RddN4 or on the Sturtevant Richmont
website.