ITEM-ITEM
QTY
SIZE
WRC/CRC – *
1
¾"
10.1.2.4 Touchpanel
Conduit
The control cable must be run to the touchpanel location with a 15' (4.5m) service loop.
Conduit required to pull the cables listed below will be determined by the Hospital/
Contractor.
Power
One (1) electrical wall outlet adjacent to the empty junction box.
Cabling
One (1) Belden 8723 or 88723 cable from the junction box to the controller (maximum
300' [91.5m])
10.1.2.5
All Displays
Conduit
The signal cable must be run from the system controller to a double gang junction
box with single gang mud ring flush mounted in the wall directly above the mount-
ing bracket with a 15' (4.5m) service loop. Conduit required to accomplish this will be
determined by the Hospital/Contractor.
Structural
Hospital/Contractor to mount the Stryker provided mounting bracket to the wall (for
40" LCD option) in the desired location with proper reinforcement to support the cor-
rect display before Stryker installation. The mounting bracket and instructions will be
delivered with the LCD. (See Appendix A for equipment weight, power, and dimension
specifications.)
Power
One (1) standard electrical outlet mounted adjacent to the empty junction box. (See
Appendix A for power consumption.)
Cabling
One (1) Belden 8241 or 88241 (or equivalent) from the display to the controller (maxi-
mum 1000' [305m])
10.2 SuiteStatus Systems
The SuiteStatus System combines the Fixed, IP Camera video signals from separate rooms into one
display located at the OR Control Desk. All video feeds will be sent over a provided network, using the
Internet Protocol standard, and are accessible from up to three All-In-One Computer Systems. This
section describes site preparation requirements necessary prior to the installation phase of the SuiteS-
tatus System.
Note
All networking requirements should be met prior to Stryker’s arrival for installa-
tion.