25
S
6.5.4
SD CODEC (L1)
Dimensions
19.5"W X 2"H X 12"D
Data
• One (1) Ethernet connection
• Please consult Project Manager if ISDN is utilized.
Power
One (1) standard outlet within 3'
6.6
IBC Podium with Touch Panel
LA
Note
Customer/Contractor responsible for installation of podium prior to Stryker
installation.
Dimensions
32"W X 46"H X 24.3"D
Cabling
• Cables provided by the Hospital/Contractor for runs over 50'.
• Cables pulled by the Hospital/Contractor.
• Leave a 15' (4.5m) service loop at each end for terminations.
• Cable Specifications: Cables may vary depending on conference room configu-
rations. Please consult Project Manager for site specific requirements:
• Touch Panel:
º
One (1) Belden 8723 or equivalent (maximum 1000')
º
One (1) Extron MHR5 or equivalent (maximum 150')
• Laptop Connection:
º
One (1) Belden 8723 or equivalent (maximum 1000')
º
One (1) Extron MHR5 or equivalent (maximum 150')
• PC Connection:
º
One (1) Belden 8723 or equivalent (maximum 1000')
º
One (1) Extron MHR5 or equivalent (maximum 150')
• Podium Microphone
º
One (1) Belden 8723 or equivalent (maximum 1000')
• Network
º
Two (2) CAT6 cables (maximum 328')
Conduit
Two (2) 2" conduits
Back Box
• 12"W X 12"H junction box
• Flush mounted in the floor under the podium
Power
• One (1) standard outlet
• Flush mounted in the floor beneath podium adjacent or within 12" X 12" junc-
tion box
ITEM-ITEM
QTY
SIZE
IPD – ICW
2
2"