24
S
6.4
IBC with Customer Supplied Casework (ICW3)
LA
Data
Per listed equipment
Back Box
• One(1) 18"W X 18"H X 4"D (minimum) junction box flush mounted.
• Set bottom of the box 9" above finished floor within the foot print of the tower.
• Terminate all integration conduits to this junction box.
• Hospital/Contractor will provide and pull all cables.
Power
• One (1) - 20 amp circuit for outlets behind video router.
• One (1) - 20 amp circuit for quad outlet behind casework.
Space
Requirements
• Must allow for a minimum 2" cable passage between all components housed
inside.
• Section housing video router must have an interior dimension of at least
27.5"W X 31"H X 29"D.
• Section housing the video router must be vented.
• Must allow for direct access to backboxes per requirements listed below.
Note
Customer/Contractor responsible for installation of casework prior to Stryker
installation.
6.5 Casework Equipment
6.5.1
SwitchPoint Infinity 2 (A2)
LA
Dimensions
• Router 19.5"W X 28.5"H X 23"D
• Total Space Required: 27.5"W X 31"H X 29"D
Data
One (1) Ethernet connection
Back Box
• One(1) 18"W X 18"H X 4"D (minimum) junction box flush mounted.
• Set bottom of the box 9" above finished floor.
• Terminate all integration conduits to this junction box.
Power
One (1) standard outlet within 3'.
6.5.2
SPI2 Touch Panel, 19" (B2)
Dimensions
16.9"W X 15.5"H X 8.2"D with stand
Power
One (1) standard outlet within 3'
6.5.3
Wireless Microphone Receiver (M)
Dimensions
8.3"W X 1.93"H X 7.2"D
Power
One (1) standard outlet within 3'