32
S
Back Box
• 4"W X 4"H junction box with dual-gang mud ring
• Flush mounted in wall 12" below finished ceiling (wall mounted only)
• Flush mounted in ceiling (ceiling mounted only)
Power
None
ITEM-ITEM
QTY
SIZE
3WP/3CP – A (SPI3)
1
1"
IWP/ ICP – ICW (IBC)
1
1"
8.9 Wall/Ceiling Mounted HD Pan/Tilt/Zoom Camera
LA
Conduit
One (1) 1" conduit
Cabling
• Cables pulled by the Hospital/Contractor
• Leave a 15' (4.5m) service loop at each end for terminations
• Cables provided by Stryker (max length 50')
Back Box
• 4"W X 4"H junction box with single-gang mud ring
• Flush mounted in wall 12" below finished ceiling (wall mounted only)
• Flush mounted in ceiling (ceiling mounted only)
Power
None
ITEM-ITEM
QTY
SIZE
3HW/3HC – A
1
1"
8.10 Echo Canceling Microphone with Internal Speaker
LA
Conduit
One (1) 1" conduit
Cabling
• Cables provided by the Hospital/Contractor for runs over 50'.
• Cables pulled by the Hospital/Contractor.
• Leave a 15' (4.5m) service loop at each end for terminations.
• Cable Specifications:
º
Four (4) Belden 8723 or equivalent (maximum 1000')
Back Box
• 4"W X 4"H junction box with single-gang mud ring
• Mounted in desired location near microphone
Power
One (1) standard outlet within 18" of touch panel location.
ITEM-ITEM
QTY
SIZE
3EM – A (SPI3)
1
1"
IEM – ICW (IBC)
1
1"
LEM – S (Level)
1
1"