![Stryker LUCAS 3 Скачать руководство пользователя страница 56](http://html.mh-extra.com/html/stryker/lucas-3/lucas-3_service-manual_1385958056.webp)
LUCAS
®
3 Chest Compression System –
Service Manual
3328798-002, ©2017 Physio-Control, Inc.
Page
56
of
91
Preventive Maintenance
Maintenance Procedure
NOTE:
Use PIP checklist 3326789 Rev C to record PIP results during the maintenance procedure.
Step
Routine
Description
Ref. /
1
Cleaning
Clean the device according to IFU, replace if
necessary
6.1 in the
IFU
2
Check/Replacement Bellows
Check Bellows is intact and clean, replace if
necessary according to
3
Check/Replacement Suction Cup
Check Suction Cup is intact and clean, replace if
necessary
6.2 in the
IFU
4
Check/Replacement Patient
Straps
All fabrics (Patient Straps, Stabilization Strap and the
Carrying Bag) are checked with the aspect of
cleanliness and that the Velcro and buckles isn’t worn
and fulfils its function. Replace if necessary.
6.3 in the
IFU
6.4 in the
IFU
5
Check/Replacement Stabilization
Strap
6
Check/Replacement Carrying
Bag
7
Check/Update Software
Check the software version in the device and update if
possible according to
8
Check for Errors and Warnings
Start the device and let it run for a minute to check
that no errors or warnings occur according to
9
Clean Fan and Mesh Grill
According to
Clean Electric Fan and Mesh Grill
10
Compression Module Lubrication
According to
Compression Module Lubrication
11
Claw lock lubrication
According to
12
Function Check
Perform a function check according to