LUCAS
®
3 Chest Compression System –
Service Manual
3328798-002, ©2017 Physio-Control, Inc.
Page
37
of
91
ongoing compressions.
Category (continued)
Specifications (continued)
Suction Cup Start Position
Factory default setting
QuickFit: The operator manually lowers the Suction Cup to the chest. When
pushing the PAUSE key, coming from ADJUST mode, the LUCAS device fine-
tunes the Suction Cup height position to the chest within a distance of 1.2 inches /
30 mm, and then the LUCAS device locks the Start Position.
Setup options
The device can be setup for QuickFit, AutoFit or Manual.
AutoFit: The device automatically lowers the Suction Cup from its upper position
down to the chest and finds and locks the Start Position. The device will do the
AutoFit when the operator pushes PAUSE key coming from ADJUST mode.
Manual: The operator manually lowers the Suction Cup to the chest. When
pushing the PAUSE key, the LUCAS device locks the Start Position. No fine-tuning
will occur.
Suction Cup in ADJUST
mode
Factory default setting
Manual: The Suction Cup has to be pulled up manually
Setup options
The device can be setup so that the Suction Cup automatically returns up from the
chest when the operator pushes the ADJUST key coming from PAUSE or ACTIVE
modes.
Pressure pad in PAUSE
mode
Factory default setting
The device stops compressions and locks the pressure pad in its Start Position.
Setup options
To allow for chest rise during ventilation, the device can be setup so that the
pressure pad moves up 0.4 inch / 10 mm above the Start Position during PAUSE.
Pressure pad during
ventilation pauses in
ACTIVE modes
Factory default setting
The device temporarily stops compressions and locks the pressure pad in its Start
Position.
Setup options
To allow for chest rise during ventilation, the device can be setup so that the
pressure pad moves up 0.4 inch / 10 mm above the Start Position during
ventilation pauses.
Pressure pad in ACTIVE
modes
Factory default setting
The pressure pad returns to Start Position between each
compression
Setup options
To allow for chest rise during asynchronous ventilation, the device can be setup so
that the pressure pad moves up 0.4 inch / 10 mm above the Start Position at every
compression.