Make sure all parts are assembled as picture below. The screws shall be locked with thread lock fluid and
tightened to 9 in-lbs / 1.0 Nm with a Torque Screwdriver
Flat 5.5x0.8 mm (5.5).
•
Connect the Fan Cable to the “Optional” connector on the Control PCB. Place the cable between the coils
according to the picture below.
•
Put back the Hood according to
Reassembling
instruction in
3. Test
•
Perform a Function Check according to
•
When the device is in active mode let it run, after approximately 10 minutes the fan should start. Now set
the device in adjust mode, the fan should continue running. Check that there is no unusual sound from the
fan. When the temperature inside the hood has decreased to 30°C (86°F) the fan should stop. Continue
with the Function Check.
Fan Cable connection
Threaded plate
Cable outlet
Fan
Fan washer
Mesh
Bracket
Screw X4