Remove the Communication PCB
2. Reassembling
•
Put the new Communication PCB in place with the new plastic screws, plastic spacers, and the nuts
•
Finger tightens screws by holding screw head lightly with fingertip until all items are tight against each
other and no play is visible. Do not overtighten.
•
Then use tools to tighten screw and nut ¾ of turn (27
0̊), tightening order according to the numbers shown
in the picture
•
Flatten screws (3x) using a non-knurled flat nose plier right up against the nut shown in the picture below
•
Put back the Controller PCB according to
Reassembling
instruction in
3. Programming
•
Connect a laptop and program the Protective PCB according to
•
Check that “Device S/N”, Time and Date, “Total Number of Strokes”, “Hardware ID” and “Device Friendly
Name” are correct (if possible check against “Store View” performed in Disassembling step). If “Store View”
was not possible set: “Device S/N” (e.g. 35160021) and “Hardware ID” (e.g. 100921-00) according to Type
Label, “Total Number of Strokes” to 0 (zero) and “Device Friendly Name” to “LUCAS3-“ + serial number
according to Type Label.
4. Test
•
Perform a Function Check according to