Clean Electric Fan and Mesh Grill
Cover the Carrier Ball Screw hole with a piece of tape to keep dust out of lubrication area.
From the inside of the device, clean the Fan and Mesh Grill with compressed air. Clean dust from areas on the device
that require lubrication. Remove tape from Carrier Ball Screw hole.
Compression Module Lubrication
Use only specified Carry Ball Screw Grease (21576-000018)
With the hood off pull down the Piston and the Adjustment Tube until the Carry Ball Screw is visible in the Lubrication
Hole.
Apply the grease onto the Carry Ball Screw with the syringe (0.5 ml).
Move the piston manually up and down 5 times over the full range to work the grease in.
Tape
Carry Ball Screw