INT_POTHALL
No=0x37
PotPos./Hall
sensor mismatch
after a correct
start.
Control POT fault or
un- linear. A start with
13N release. Rough
compression
movement.
Preform a new
POT calibration.
If warning
repeated replace
the compression
module.
LUCAS V2
operation.
INT_CURRENT
No=0x38
Current fault
Mean current to motor
>20A. Jam of
compression module.
Current read fault.
Motor.
Compression
module. Control
board.
None.
INT_P_REBOOT
No=0x39
Protective reboot
in active mode
Protective electrical
fault.
Protective board.
Restart of
startup test.
INT_1msCODE
No=0x3A
Warning Access of
1ms Code
timeout.
Halted or disturbed
Control CPU.
Control board.
None.
INT_C_HALL
No=0x3C
Motor Hall sensor
fault.
Motor or Hall sensor
cable fault.
Check Hall
sensor cable or
replace Motor.
Slow or weak
compression
s.
RUN_ADJUST
No=0x46
Incorrect Motor
power at adjust
mode
Too high Motor Power
(>25W) at Adjust
servo.
Incorrect current read.
Motor. Control
board.
Adjust servo
switched off.
RUN_13N
No=0x49
13N release (force
UP too high)
13N release > 10mm.
Too high start position
or mechanical locked
compression
movement.
Assure correct
start and
movement of the
compression
module. If
needed replace
the compression
module.
13N release
information
signal (3 fast
signals).
Start
Position
adjustment
required
COM_P_TIMEOUT
No=0x50
I2C Protective
Communication
Timeout
No I2C communication
with Protective system
>100s. I2C
communication fault or
halted Protective CPU.
Check cable
between boards.
Restart LUCAS3
and run for 30
minutes. If fault
repeats replace
Protective board.
Device
cannot be
started in
active mode.
COM_RTC
No=0x53
Real Time Clock
Fault
Reel Time Clock read
fault at start. SPI com
fault or too low “BAT1”
voltage.
Charge “BAT1”
by starting the
LUCAS3 and
press PAUSE
leave the
machine ON for
30 minutes, set
correct time and
date with the
LUCAS3.exe
program. After
this restart and
check for
COM_RTC
warning. If
needed replace
the Control
board.
LOG file date
and time
fault.
Hot restart
after battery
switch not
functional.
Internal fault
- contact
Technical
Support
COM_SD_MEM
No=0x54
Fault SD card
Communication
SD card not inserted
or broken.
SD card.
No LOG file
created.