![Stryker LUCAS 3 Скачать руководство пользователя страница 69](http://html.mh-extra.com/html/stryker/lucas-3/lucas-3_service-manual_1385958069.webp)
LUCAS
®
3 Chest Compression System –
Service Manual
3328798-002, ©2017 Physio-Control, Inc.
Page
69
of
91
•
Put back the Drive Belt according to
Reassembling
•
Put back the Compression Module according to
Reassembling
•
Put back the Hood according to
Reassembling
instruction in
•
Put back the Bellows according to
Reassembling
•
Put back the Suction Cup.
3. Test
Perform a Function Check according to
How to Replace the Control PCBA
1. Disassembling
•
Before disassembling and changing any PCB and if possible; connect a laptop and perform “Store View”
using
•
Follow
Disassembling
instructions in
How to Replace the Compression Module
Compression Module.
•
Remove the two remaining screws from beneath and lift out the PCB carefully.
•
Warning
: Always use ESD protection when handling PCBs!
•
Remove SD-card from old Controller PCB. Check that the SD-card is intact and readable/writeable. If SD-
card is working properly keep it for the new Controller PCB.
•
Remove Communication PCB according to the
Disassembling
Terminal
Motor Cable Outlet