LUCAS
®
3 Chest Compression System –
Service Manual
3328798-002, ©2017 Physio-Control, Inc.
Page
88
of
91
Electronics Test
Turn on the device and perform the following tests:
1. Check that the Internal Function Test is performed and that the ADJUST LED shows a green light.
2. Change mode to ACTIVE (30:2) and check that the ACTIVE (30:2) LED shows a green light. Let the device run
for approximately two minutes and listen for unusual noise. Check that there is an audible alert sound prior to the
ventilation pause (each 30 strokes at default settings) together with an intermittent LED.
3. Change mode to PAUSE and check that the PAUSE LED shows a green light.
4. Change mode to ACTIVE (continuous) and check that the ACTIVE (continuous) LED shows a green light. The
LED will then blink each ten strokes (ventilation alert).
5. Change mode to ADJUST, pull the Suction Cup down to test the Adjustment Servo. Push the Suction Cup back
up to its top position. Check that it runs smoothly.
6. Connect a charger to the device and check that the charging sequence begins (the battery LED’s show a
“running” light). Disconnect the charger.
7. Turn the device off.
8. Testing Transmit mode by following
Connect the device via Bluetooth
to make connect to a PC using LUCAS
®
3
Program Loader.
•
If the device has undergone Maintenance and has passed the Function Check press “Set PC Date” to change
“Last Service Date”.
•
If the device has undergone a Repair without Maintenance, do NOT press “Set PC Date” (a connection
between the device and laptop still has to be established to check that the communication is working).
•
Press
OFF
in LUCAS
®
3 Program Loader to switch off the Bluetooth communication. Check the Bluetooth
light is off.