Alarm conditions
Remove from use mode
The Remove From Use (RFU) is a safety mode to limit operations. A fault condition prevents the product from normal
functions and requires service. The controller will stop the Active Therapy and communicate to the operator that the
controller is going into RFU mode.
CAUTION
Always remove the product from use before servicing any components. Contact qualified service personnel for service.
Depending on the remove from use (RFU) condition, text may or may not be displayed. For example, if there is a power
loss.
•
Water temperature probes are out of the allowed range
•
Program and data checksum error
•
High thermal cutout test failed
•
Backup power product replacement required
•
Low or over safety temperature
•
Pump over current
•
Compressor power fault
•
Heater power fault
•
Refrigerant control valve fault
•
Main DC power lost
•
CAN heart beat lost
•
Dual safety temperature sensors do not match readings
•
Dual safety temperature sensors are out of the allowed range
•
Hardware watchdog heartbeat failure