Cleaning
Cleaning the external surfaces
Clean the external surfaces of the controller and system components before each use. System components may be
subject to contamination during use from contact with soiled hands of the user, airborne pathogens, and unexpected or
accidental events. Make sure you remove all visible soils.
CAUTION
Do not power wash this product.
Tools Required:
•
Mild soap
•
Soft, lint free cloth (2 or more)
Validated mild soap:
•
Enzol® Enzymatic Instrument Cleaner by Johnson & Johnson
To clean the external surfaces of the controller and system components:
See
Product illustration on page 10
for clarification of product component names and locations.
1.
Unplug the power cord from the wall outlet.
2.
Apply wheel locks.
3.
Undo power cord and hose straps.
4.
Unravel and lay out hoses, cables, and power cord.
5.
Disconnect the hoses. Push back on the retaining collar of the port on the controller. Pull the hose to disconnect.
6.
Disconnect the patient temperature probe cable from the port.
7.
Remove the reservoir. Pull forward at an angle, and lift out the reservoir.
8.
If necessary, empty the water from the reservoir. Dispose of the water per your hospital protocol.
9.
Prepare a mild soap and water solution as described by the manufacture.
10. Wipe the inside and outside of the reservoir and reservoir lid, with a soft, lint free cloth moistened with soap and
water solution.
11. Wipe the hoses and patient temperature probe cables, with a soft, lint free cloth moistened with soap and water
solution.
12. Wipe the controller surfaces while the reservoir is removed with a soft, lint free cloth moistened with soap and water
solution. Also wipe the following system components:
•
Hose connectors
•
Power cord
•
Hose and power cord management straps
•
Storage compartment door
•
Inside storage compartment
•
Graphical user interface display
•
Handle
13. Wipe the controller, reservoir and reservoir lid surfaces, and system components with a clean, dry cloth to remove
any excess liquid.
14. Replace the reservoir.
15. Allow the external surfaces of the controller and components to dry thoroughly.