53
Alarm conditions
Water temperature deviation alarm group (Continued)
Alarm generation:
1.
If the actual water temperature is 0.8° C or more above or below the Final Target Temperat ure, the product will
display the Water Temperat ure Deviation Alarm.
2. When the product is entering the Manual mode or you change Target Temperature , the product will pause the audible
component of the Water Temperature Deviation Alarm for four hours. The four hour pause automatically cancels
after the Water Temperat ure becomes equal to the Final Target Temperature .
Check water flow alarm
This alarm notifies the operator of the quality of the flow in each individual water circuit.
Alarm generation:
•
When in Manual or Automatic Mode and several ports are in use for the therapy.
•
You have selected an outlet port and the flow is lower than 0.8 l /min for a period of 60 seconds or more. The
product will display a Check Water Flow Alarm for the given port.
Notes
•
The alarm displays if the flow is not at an optimal level on each port. This alarm will ask the operator to confirm
which ports are currently in use.
•
The addition of a port does not need the operator to confirm the addition.
•
The removal of a port requires the operator to confirm.
•
The Check Water Flow Alarm for the given outlet port stops, if the operator confirms removal.
•
When none of 3 ports has a flow higher than or equal to 0.6l / min, the product deactivates the heater exchange and
generates a no flow alarm. Otherwise the heater exchange will remain active as indicated by the current mode.
Power backup level alarm
This alarm notifies to the operator an indicator of the status of the Backup Power Level.
Notes
•
The indicator will remain active until a qualified technician replaces the battery.
•
There is no reduction in the usability of the product. The product remains functional and a visual alarm is displays.
•
The product will disable the Backup Power Level Alarm in Sleep Mode. Otherwise, the alarm is enabled.
Alarm generation:
The product will display Backup Power Level Alarm when the battery level backup power is less than 100 minutes of
alarms. Once activated, the Backup Power Level Alarm will remain active until you power off the product.
Therapy paused time out alarm
This alarm converts a therapy pause into an alarm if the duration of the pause is too long.
Alarm generation:
When paused for five minutes, the product will display the Therapy Paused Time Out Alarm. After you resume the current
therapy, the Therapy Paused Time Out Alarm is deactivated.