www.stryker.com
8001-009-001 REV G
25
Operation
Connecting and disconnecting thermal transfer devices (Continued)
Figure 14: Hose clamps
Note:
The term “thermal transfer devices” is used throughout this manual and is interchangeable with blankets and
wraps, unless indicated otherwise.
Always clamp the hoses before disconnecting. See
Draining the thermal transfer devices on page 33
Powering on the product
The operator should stand in front of the controller within arm’s reach. This allows the operator to see and respond to the
display notifications.
CAUTION
•
Shock Hazard. Improper handling of the power cord may damage the power cord and cause potential shock hazards.
If damage has occurred to the power cord, immediately remove the
Altrix
system from service to avoid the risk of
serious injury or death. Contact the appropriate maintenance personnel.
•
Electric shock. This equipment must only be connected to a supply mains with protective earth.
•
Always plug this product directly into a properly grounded hospital-grade or medical-grade wall outlet to achieve
grounding reliability.
•
Do not use high frequency surgical instruments or endocardial catheters while the
Altrix
system is in use. This is to
avoid the risk of electrical shock, burns, or electromagnetic interference.
•
Explosion risk. This product is not suitable for use in the presence of flammable anesthetic mixture with air or with
oxygen or nitrous oxide other than nasal or mask type.
•
Do not place cables, hoses, or power cord in walkways to avoid the risk of trip hazards.
•
Avoid reduction in water flow. Do not connect two or more thermal transfer devices in a series on a single port.
•
Do not use three or more adult Mul-T-Blanket’s at the same time to avoid the risk of water overflow when you power
off the controller.
•
When you operate the product near ambient temperature limitations of 15.0° C (59.0° F) or 32.0° C (89.6° F), you
may experience a reduction in product performance.
To start the product:
1.
Plug the power cord into a wall outlet.
2.
Tap the Stand-by button to start the product.
3.
If you are going into Automatic mode or Monitor mode, see
Selecting and setting the primary probe on page 27
. If
manual mode, got to the next step.
4.
See
Removing and replacing the reservoir on page 26
5.
See
Filling the reservoir with sterile distilled water on page 26
6.
Connect up to three thermal transfer devices (with the exception of adult Mul-T-Blankets) to dedicated adapter
hoses and ports.