Setup
Unpack the cartons and check all items. Make sure that the product is free from visual damage before putting into
service.
CAUTION
•
Shock Hazard - Improper handling of the power cord may damage the power cord and cause potential shock
hazards. If damage has occurred to the power cord, immediately remove the temperature management system from
service to avoid the risk of serious injury or death. Contact the appropriate maintenance personnel.
•
Take special precautions regarding electromagnetic compatibility (EMC) when using medical electrical equipment
like
Altrix
. Install and place
Altrix
into service according to the EMC information located in the EMC section of this
manual. Portable and mobile RF communications equipment can affect the function of
Altrix
.
•
Shock Hazard. If the internal electrical components are exposed, because the side panel or cover are compromised,
remove the product from use.
•
Always make sure that the product reaches room temperature before you setup or operate the product.
•
Before first use, disinfect the internal water circuit.
Inspecting
Before you place the product into service, make sure that the controller works.
1.
Visually inspect the product for any signs of shipping damage.
2.
Plug the product into a properly grounded, hospital grade wall receptacle. Make sure that the power indicator
illuminates on the operator control panel.
3.
Before first use, .
Disinfect the internal water circuit and hoses every 14 days on page 39
Selecting a language
The
Altrix
controller has several language choices. English is the default language.
To select a language when in standby mode:
1.
Tap the Settings button, to show the Select Language screen.
a.
Tap Next, if you are in therapy mode.
2.
Tap More to view other languages.
3.
Select a language. Tap on the Increase or Decrease buttons or tap the language, to highlight a language of your
choice.
4.
Tap Confirm.
Note:
If you do not touch the screen for three minutes, the LCD will return to the previous menu.
Testing visual and audible alarms
Before placing the product into service, make sure that the visual and audible alarms are functioning.
1.
Tap the Settings button.
2.
Tap the Back button.
3.
Tap the Visual/Audible icon.