www.stryker.com
8001-009-001 REV G
15
Introduction
Product functions (Continued)
Icon
Name
M
ed
Medium: patient temperature increases at a rate of 4.0° C in 12 hours (0.33° C/ hour).
M
ax
Maximum: water temperature approaches water target as fast as possible
Min
Minimum: patient temperature increases at a rate of 4.0° C in 24 hours (0.17° C/hour).
S
et Custo m
Set Custom: patient temperature increases at a customized temperature and time period the
operator selects. The temperature increases 0.05° C/hour to 0.5° C/hour
Max
Maximum: water temperature approaches water target as fast as possible
Med
Medium: water is cooled to target, with a max of 15.0° C difference between the patient and
the water temperature
Min
Minimum: water is cooled to target, with a max of 10.0° C difference between the patient and
the water temperature
Product alarms
Audible alarms work in conjunction with the display.
Alarm priority and description
Priority alarm
Audible reminder
Icon flashes
Medium
Repetitive burst of three beeps every 25
seconds
When a medium priority alarms, the icon will
flash to indicate there is an alarm. It will
continue to flash until the alarm is resolved.
Low
Single burst of two beeps
The icon does not flash when there is a low
priority alarm.
Audio pause
Button breathes as a reminder
Pausing the alarm will not stop the icon from
flashing.
Note:
You can pause the audio alarm. The alarm will resume within five to ten minutes or sooner, if not resolved. The
alarm will resume depending on when the alarms became active and the number of active alarms.