48
8001-009-001 REV G
www.stryker.com
Troubleshooting
Problem
Possible cause
Action
Remove from use
Patient temperature
Out of range
Check the placement of the
probe
Not applicable
Patient temperature output
(PTO)
External device output
displays high value >45° C
when input is out of range
(As a result of one of the
following: patient probe
disconnected, controller in
standby / sleep mode,
patient temperature outside
the range of 25° - 45°C).
To resume calibration:
•
Disconnect the
external device from
the reusable adapter
cable
•
Tap the Help button to
display the alarm
screens
•
Find the “Temperatu re
Output Alarm” screen
•
Tap Confirm to restart
calibration
•
Wait until the “Monitor”
is solid
•
Connect the external
device to the reusable
adapter cable
Not applicable