8
8001-009-001 REV G
www.stryker.com
Introduction
Indications for use (Continued)
•
For use in all clinical settings including coronary care units, operating, recovery and emergency departments, burn
units, and medical/surgical units
•
Adult and pediatric patients
•
Monitoring and controlling patient temperature
•
Temperature reduction in patients where clinically indicated, e.g. in hyperthermic patients
Intended users
•
Physicians
•
Advanced Practice Registered Nurses
•
Nurses
Expected service life
The
Altrix
controller has a five year expected service life under normal use, conditions, and with appropriate periodic
maintenance. See the maintenance manual for preventive maintenance and service information.
Contraindications
For core body temperature regulation:
•
Raynaud’s Phenomenon (primary or secondary)
•
Application to lower extremities distal to aortic cross-clamping
Specifications
Model
8001-000-001
Electrical Requirements - AC Voltage Input Current and Voltage
Ratings
120VAC, 60Hz 12A
Physical dimensions
Height
42.5 in.
107.9 cm
Width
15.0 in.
38.1 cm
Depth
23.0 in.
58.4 cm
Empty weight
150.0 lb
68.0 kg
Filled weight
160.5 lb
72.8 kg
Reservoir capacity
1.3 gal
5.0 L
Water temperature
Control setting range
39.2° - 104.0° F
4.0° - 40.0° C
Control accuracy
±0.3° C (4.0° - 40.0° C)
Display measurement accuracy
±0.2° C (4.0° - 40.0° C)
Display / resolution setting
0.1° C
Default setting
104.0° F
40.0° C