40
8001-009-001 REV G
www.stryker.com
Disinfecting
Draining the internal water circuit and hoses for disinfection (Continued)
Note:
For best results, the floor drain should be within reach of a wall outlet to power on the controller.
3.
To drain the controller, pull up on the controller drain plug (A) to open the drain (
). Leave the
drain open.
A
Figure 24: Drain plug
4.
Connect a hose to each port (
Figure 25: Hoses connected
5.
Close the connector ends of all three hoses:
a.
If you have
Colder
style connector hoses, attach the service tool adapter hose (8001-999-017) (
). Complete this for all three hoses.
+
=
Figure 26: Colder style connector hose connected to a tool adapter hose
b.
If you have
Clik-Tite
hoses, make sure that the connector ends are connected and closed (A), and clamps are
open (B). Complete this for all three hoses.